1.1 This test method covers a procedure for the fatigue testing of metallic tibial trays used in partial knee joint replacements. 1.2 This test method covers the procedures for the performance of fatigue tests on metallic tibial components using a cyclic, constant-amplitude force. It applies to tibial trays which cover either the medial or the lateral plateau of the tibia. 1.3 This test method may require modifications to accommodate other tibial tray designs. 1.4 This test method is intended to provide useful, consistent, and reproducible information about the fatigue performance of metallic tibial trays with unsupported mid-section of the condyle. 1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.6This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.1.7This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Product Details
Published: 09/01/2023 Number of Pages: 6 File Size: 1 file , 390 KB Redline File Size: 2 files , 560 KB Note: This product is unavailable in Russia, Ukraine, Belarus